Transmitter |
Housing |
Al wall hosing robust design protection IP65
240¡¿240¡¿120mm,approximately 5.5kg |
Mounting |
4 holding flaps H¡¿B=142.5¡¿273 for ¥Õ5 |
Power
supply |
AC90~270V,45~68Hz:Max.45 VA
Alternative DC:18~36V
DC max. 1000mA for 1 to 2 sensors
DC max. 1800mA for 3 to 4 sensors |
Display |
Monochrome-L/C 4/1-VGA,320¡¿240 pixel
Illuminated,Large display of the measurement
result and drag pointer with analogue bar graph
displays of deviation of the measurement in its
defined scale |
Handing |
Foil keyboard with 4 soft keys
Well-structured multilingual menu |
Status
display Outputs |
3 LEDs for operational signal, of the measurement
and Access-PIN
for parameter Outputs 4¡¿0/4~20mA Isolated outputs for concentration
range can be seet for each each channel specific Load max. 500¥Ø
|
Data
interface |
RS232,9600Bd for data communication or comfortable
software-update |
Data
memory |
EEprom-data sets.All data are absolutely saved
in case of power failure
Option:additional changeable memory module for back-up of 30 data
sets
|
Input |
4¡¿TNC-plugs for 50¥Ø coaxial cable, RG58,
150m for simultaneous running of 4 separated microwave modules.
Digital transmission on
carrier frequency 10MHz |
Microwave
module ¥ì-ICC 2.45 |
Housing |
Al wall hosing robust design protection IP65
166¡¿100¡¿81mm,approximately 1.4kg |
Mounting |
4 holding flaps H¡¿B=142.5¡¿273 for ¥Õ5 |
Power
supply |
DC24V through coaxial cable from controller |
Microwave |
extremely stable phase and attenuation measurement
with PLL=synthesiser :2.45GHz Transmission performance 10mW, 10dBm |
Microwave
cable |
2¡¿N-plugs for coaxial cable 50¥Ø typical 1 to
2m (max. 3) |
Data
memory |
Eeprom-data memory for reference point
in case of power failure all data absolutely saved |
Output |
1¡¿TNC-plugs for coaxial cable 50W, RG58 for
data transmission to the controller |
Input |
PT100/PT1000 two-wire connection Section-50
to 200¡É |
Temperature |
Environment 0 to 60¡É |